Summary of Test Results for a Robustness Test in Accordance with part 3 of the DIN EN 14175 series of standards

In reference to the standardised testing procedure in accordance with part 3 of the DIN EN 14175 series of standards, the retention capacity of a WALDNER Scala RWI-TA 1500 × 900 laboratory fume hood was examined with and without the integration of the SmartRack® and SmartFrame.

Summary of Test Results for a Robustness Test in Accordance with part 3 of the DIN EN 14175 series of standards

Test Procedure:

In reference to the standardised testing procedure in accordance with part 3 of the DIN EN 14175 series of standards, the retention capacity of a WALDNER Scala RWI-TA 1500 × 900 laboratory fume hood was examined with and without the integration of the SmartRack® and SmartFrame. Five measurement runs were carried out in each case and subsequently averaged.

Results:

  • The average peak concentrations were 0.86 ppm without SmartRack® and 1.23 ppm with SmartRack® (limit value 3.25 ppm according to part 3 of the DIN EN 14175 series of standards).
  • The average values were 0.14 ppm without SmartRack® and 0.18 ppm with SmartRack® (limit value 0.65 ppm according to part 3 of the DIN EN 14175 series of standards).
  • Both values are therefore well below the permissible limit values.

     

     

Grafik Abzugsleistung EN
Figure: Comparison of the test gas concentration (c) in a WALDNER Scala series fume hood with rear wall installation (RWI-TA) 1500 × 900 – 900 with WALDNER Airflow Controller (AC), shown without SmartRack® and SmartFrame (dark green) and with SmartRack® and SmartFrame (light green).
Robustheitstests nach EN 14175 für SmartRack und SmartIntegrate
Figure: SmartFrame WA 002 (item number 04010) and SmartRack® classic (item number 01704) in a WALDNER Scala 1500 RWI.

Technical Evaluation

The minimal changes in test gas concentration caused by the SmartRack® and SmartFrame have no significant impact on the safety-relevant function of the fume hood. The extraction performance in accordance with part 3 of the DIN EN 14175 series of standards is fully maintained and remains safe.

Conclusion

Even when SmartRack® and SmartFrame are used simultaneously in a fully equipped WALDNER laboratory fume hood (including glassware and laboratory equipment), the limit values of the robustness test in accordance with part 3 of the DIN EN 14175 series of standards are complied.

The use of the SmartRack® and SmartFrame in the WALDNER Scala RWI-TA 1500 × 900 laboratory fume hood fulfils the requirements of the robustness test in accordance with part 3 of the DIN EN 14175 series of standards and does not present any significant deterioration in air technology.

Download our flyer on this topic here

 

Summary of test results for a robustness test in accordance with part 3 of the DIN EN 14175 series of standards

More Information

Here you can find more information about the concept and products of SmartIntegrate.

If you are interested in our products, please use the inquiry form in the downloads section.

If you need more information, please contact us and use our contact form.